ATLAS galda Class 1 biodrošības kabineti
NEOTERIC
Work Chamber
- hinged front sash made of laminated safety glass; opening, closing and upholding is carried out by gas springs;
- damper against front sash closing impact;
- side windows made of tempered glass;
- slanted front surface of the cabinet;
- work chamber lighting;
- two electric sockets in the work chamber (in the right part of the rear wall);
- airflow laminarization screen made of polymer micromesh;
- three-section tabletop made of 1.5 mm mirror stainless steel (AISI 304);
- work chamber tray made of 0.8 mm stainless steel (AISI 304);
- tabletop reinforcements (3 pcs.);
- tabletop supports (2 pcs.);
- removable armrest made of 0.8 mm stainless steel (AISI 304)
UV Unit
- pullout (when inoperative, is stored outside of the work chamber under the tabletop);
- metal grille protecting the UV lamp;
- dampers against any impact upon lowering the work opening screen.
Supply and Exhaust Air Filtration System
- supply air coming into the work chamber is cleaned by supply HEPA filter H14;
- exhaust air coming out of the cabinet is cleaned by exhaust HEPA filter H14;
- one fan ensuring airflows inside the cabinet
Control System Components
- microprocessor control system;
- pushbutton control panel with LCD screen;
- electromagnetic reader and electromagnetic access keys;
- airflow sensors;
- front sash and UV unit position sensors;
- individual fuse for power supply to the main functions of the cabinet (fan, lighting, UV lamp);
- individual fuse for electric sockets of the work chamber;
- removable power cord with holder
Testing and Adjusting Components
- exhaust opening balancing baffle for airflow adjustment;
- built-in outlet for sampling the air upstream of HEPA filters for their integrity testing.
Stand
- frame stand with footrest;
- caster wheels for cabinet relocation;
- screw supports for cabinet fixation at operating site.
MAIN CHARACTERISTICS
Installation work chamber air cleanliness class for suspended particle (aerosol) concentration as per ISO 14644-1 -for particles of 0.5µm and more -for particles of 5.0µm and more |
ISO 5 ISO M (20; ≥5µm); LSAPC |
Class of the cabinet as per EN 12469, NSF/ANSI 49 | II |
Type of the cabinet as per NSF/ANSI 49 | A2 |
Class of HEPA filters as per EN 1822-1 | H14 |
Average velocity of the inflow though the working opening, mps | 0,47±0,03 |
Average downflow velocity in the work chamber, mps | 0,35+0,01 |
Minimum illumination of the work zone (integral value determined along the whole area of the work zone), lux | 1000 |
Air recirculation, % | ≈70 |
Maximum noise level at 1 m distance from the cabinet | 59* |
* level of noise in free sound field over the sound-reflecting surface (noise level in real operating environment depends on the dimensions of the operating site as well as on the total background noise and may vary by 3-4 dB(A))
Article | 2Е-В.001-12 |
Maximum dimensions of the cabinet /WхDхH/* (dimensions without stand are in parenthesis) , mm | 1200x770x2150 |
Maximum dimensions of the stand /WхDхH/*, mm | 1195х720х670 |
Dimensions of the cabinet assembled with stand and exhaust hood** /WхDхH/,mm | 1200х770х2355 |
Maximum net weight of the cabinet assembled with stand, kg | 235 |
Maximum input power without built-in electric sockets, W | 180/110** |
Maximum allowed load on the built-in electric sockets, W | 1000 |
Clean air inflow capacity, m3/h | 795-817 |
Exhaust air capacity, m3/h | 333-378 |
* dimensions do not account for outstanding supports
* Exhaust hood is not part of the standard delivery set; can be supplied as per customer’s additional request.